| Type of Product: | Device |
| Medsafe Reference: | 36728 |
| Brand Name: | Chromphare surgical lights |
| Model: | Devices manufactured with Soffit Rings |
| Affected: | Part number - CH00000001 |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Berchtold GmbH & Co KG |
| Issue: | The manufacturer has discovered that certain lots of the soffit ring may not adequately support the weight of the ceiling cover. As a result, the cover may shift or slide down the ceiling tube. Movement of the ceiling cover can expose the area above the ceiling, which may allow particles, such as dust, debris, residue, or other contaminants, to enter the sterile field. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/4/2026 |