Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 36714 |
| Brand Name: | Aurous Centimeter Vessel Sizing Catheter |
| Model: | CK-N5.0-35-100-P-10S-PIG-CSC-20 (CATHETER SIZING 5FR PIGTAIL NON BT) |
| Affected: | 16752732, 16813575, 16839886, 16858458, 16861720 |
| Software version: | |
| Recalling Organisation: | Obex Medical Ltd Level 3, 109 Carlton Gore Road Newmarket Auckland 1023 |
| Contact Information: | 0800 656 539 |
| Manufacturer: | Cook Incorporated |
| Issue: | This issue was identified internally during production. Following issue detection, product samples underwent tensile testing, where one sample cracked while being manipulated to load into the test fixture. While there is a quality control check in place to detect this failure mode prior to product release, all tested samples underwent this quality control check prior to being tested. This is indicative that the subject marker bands may fail at a point past the quality control check. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 7/4/2026 |