Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:36714
Brand Name:Aurous Centimeter Vessel Sizing Catheter
Model:CK-N5.0-35-100-P-10S-PIG-CSC-20 (CATHETER SIZING 5FR PIGTAIL NON BT)
Affected:16752732, 16813575, 16839886, 16858458, 16861720
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Cook Incorporated
Issue:This issue was identified internally during production. Following issue detection, product samples underwent tensile testing, where one sample cracked while being manipulated to load into the test fixture. While there is a quality control check in place to detect this failure mode prior to product release, all tested samples underwent this quality control check prior to being tested. This is indicative that the subject marker bands may fail at a point past the quality control check.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/4/2026