Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 36708 |
| Brand Name: | OXOID™ Agglutinating Sera, Salmonella 9-0 |
| Model: | |
| Affected: | Lot - 6196409, 6233858, 6233859, 6260215, 6266130 |
| Software version: | |
| Recalling Organisation: | Thermo Fisher Scientific New Zealand Ltd 244 Bush Road Albany AUCKLAND |
| Contact Information: | 0800 636 327 |
| Manufacturer: | Remel Europe Limited |
| Issue: | An internal technical investigation has identified that certain lots of OXOID Agglutinating Sera, Salmonella 9-O, do not perform to Instructions for Use (IFU) criteria. Specifically, the affected lots show cross-reactivity with Salmonella 2-O within 60 seconds, which may lead to incorrect serogroup identification. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 1/4/2026 |