Medsafe Reference:36701
Brand Name:ARTIS pheno and ARTIS icono systems
Model:Catalogue numbers - 11328100, 10849000, 11327700, 11327600
Affected:ARTIS pheno, ARTIS icono with VE30B and VE40B
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue:The manufacturer has identified that the ARTIS pheno, ARTIS icono with VE30B and VE40B – Incorrectly Reported DAP value. During examination, the system may show sporadically:
1. that zero dose is administered to the patient. This sporadic behavior is only shown in fluoroscopy and/or acquisitions with one (single) frame. The dose reported is "0", but dose is actually administered to the patient. OR
2. that a major increase in the dose administered to the patient. The dose reported is significantly higher than the dose actually administered to the patient.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/3/2026