Detail of Recall

Type of Product:Device
Medsafe Reference:36692
Brand Name:TRIOS 6 System (L1P-2)
Model:TRIOS 6 intraoral scanner
Affected:1ld2522l01055b, 1ld2535l01022b
Software version:
Recalling Organisation:Osstem New Zealand Limited
10/6 Omega Street
Rosedale
Auckland 0632
Contact Information:
Manufacturer:3Shape Trios A/S
Issue:The manufacturer has identified that a reflective surface inside the TRIOS 6 scanner front tube (frame 10032019) can lead to infrared (IR)
straylight, which may produce image artefacts when using the SaMD TRIOS Dx Plus (proximal caries visualisation).
Artefacts may appear across multiple teeth in severe cases, or as localised glare in moderate cases. Localised glare may increase the likelihood of false positives for proximal caries, however, the risk of this problem leading to patient harm is low/negligible, as TRIOS Dx Plus is to be used only as an adjunct/aid. Direct visual and tactile examination performed during the checkup would likely confirm the false positive. If uncertainty remained after the clinician's examination, a confirmatory X-ray would be taken
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/3/2026