| Type of Product: | Device |
| Medsafe Reference: | 36681 |
| Brand Name: | YSIO X.pree VA20G with X.wi-D detectors digital stationary Xray |
| Model: | YSIO X.pree - with software version VA20G and X.wi-D detectors |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare GmbH |
| Issue: | In rare cases and under specific acquisition conditions involving X.wi-D detectors, the system may become unresponsive, most commonly during patient examinations that include free exposures. A system restart is required to restore operations. In some cases, the image acquired immediately before the system freeze may not be recoverable, resulting in an image loss and the necessity to repeat the exposure. Image acquisitions with MAX wi-D detectors are not affected by this problem. The manufacturer is preparing a software update (VA20H) for the YSIO X.pree system that will resolve this problem. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 24/3/2026 |