Detail of Recall

Type of Product:Device
Medsafe Reference:36681
Brand Name:YSIO X.pree VA20G with X.wi-D detectors digital stationary Xray
Model:YSIO X.pree - with software version VA20G and X.wi-D detectors
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue:In rare cases and under specific acquisition conditions involving X.wi-D detectors, the system may become unresponsive, most commonly during patient examinations that include free exposures. A system restart is required to restore operations. In some cases, the image acquired immediately before the system freeze may not be recoverable, resulting in an image loss and the necessity to repeat the exposure. Image acquisitions with MAX wi-D detectors are not affected by this problem. The manufacturer is preparing a software update (VA20H) for the YSIO X.pree system that will resolve this problem.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/3/2026