Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:36675
Brand Name: Spectral CT, IQon Spectral CT, Brilliance iCT, Brilliannce iCT Upgrade
Model:Spectral CT
Affected:Product code 728333
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Brand Name:Philips Spectral CT 7500 Systems
Model:Philips Spectral CT 7500 systems
Affected:Product code 728333
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The affected cables supply power from the Power Supply Unit to the DCDC board and then to the Detector Measurement System. Due to insufficient crimping, the cable connections can develop increased electrical resistance, which may lead to system failures and render the unit unusable until repairs are completed. During a scan, this issue may cause the scan to stop unexpectedly, generate the error code S_DMC_MODULE_VOLTAGE_OUT_OF_SPEC, or produce degraded image quality or image artifacts. In some cases, a patient re-scan may be required.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/3/2026