Detail of Recall

Type of Product:Device
Medsafe Reference:36673
Brand Name:Access Total T4 Calibrators
Model:
Affected:Lots - 538729, 538815, 538898, 539107, 539243
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The manufacturer has determined that the Access Total T4 Calibrator may produce erroneously low results on UniCel DxI 600 and UniCel DxI 800 Access Immunoassay Analyzers. Internal testing has determined that calibration curves generated using the Access Total T4 Calibrator may produce quality control and patient results approximately 23% lower than expected on UniCel DxI 600 and UniCel DxI 800 Access Immunoassay Analyzers. Falsely decreased Total T4 results reported to the physician may result in patient misdiagnosis and inappropriate treatment.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/3/2026