Detail of Recall

Type of Product:Device
Medsafe Reference:36667
Brand Name:Lutronic - CLARITY II & CLARITY II LPA
Model:LARITY II & CLARITY II LPA
Affected:Serial number - C3025C065
Software version:
Recalling Organisation:Ebos Medical Devices NZ limited
108 Wright Road
Addington
Christchurch 8024
Contact Information:
Manufacturer:Lutronic Corporation
Issue:The manufacturer has identified a potential safety-related issue may occur when operating the CLARITY II ICD system under high-fluence conditions (= 90-100 J/cm2, 1064nm) with cryogen cooling. Under these specific operating conditions, spark or loud noise (pop sound) may be observed during treatment and may increase the likelihood of localized minor burn/blister on the skin.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/3/2026