Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 36666 |
| Brand Name: | Mako Sawing power tool attachment, reciprocating |
| Model: | MICS3 Angled Sagittal Saw Attachment - catalogue number - 210490 |
| Affected: | All lots |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Mako Surgical Corp |
| Issue: | The manufacturer has identified a potential problem with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment. There is potential for the screws to loosen and fall out of the attachment during use, which may lead to a foreign object in patient. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/3/2026 |