Detail of Recall

Type of Product:Device
Medsafe Reference:36666
Brand Name:Mako Sawing power tool attachment, reciprocating
Model:MICS3 Angled Sagittal Saw Attachment - catalogue number - 210490
Affected:All lots
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Mako Surgical Corp
Issue:The manufacturer has identified a potential problem with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment. There is potential for the screws to loosen and fall out of the attachment during use, which may lead to a foreign object in patient.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/3/2026