Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 36656 |
| Brand Name: | Vantage Frame |
| Model: | Leksell Vantage Arc System |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Instrument AB |
| Issue: | The Leksell Vantage Arc System is part of the Leksell Vantage Stereotactic System used for stereotactic neurosurgical interventions. The Leksell Vantage Arc system features Arc Slide Left and Arc Slide Right. Each of which includes a locking mechanism to set the Z-Scale coordinates. The manufacturer has identified cases where the Z-Scale locking mechanism may not secure properly to the on the Leksell Vantage Arc system. Investigations confirm that this problem is associated with excessive wear on the Z-Scale locking mechanism surface. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/3/2026 |