Detail of Recall

Type of Product:Device
Medsafe Reference:36656
Brand Name:Vantage Frame
Model:Leksell Vantage Arc System
Affected:
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Instrument AB
Issue:The Leksell Vantage Arc System is part of the Leksell Vantage Stereotactic System used for stereotactic neurosurgical interventions. The Leksell Vantage Arc system features Arc Slide Left and Arc Slide Right. Each of which includes a locking mechanism to set the Z-Scale coordinates. The manufacturer has identified cases where the Z-Scale locking mechanism may not secure properly to the on the Leksell Vantage Arc system. Investigations confirm that this problem is associated with excessive wear on the Z-Scale locking mechanism surface.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/3/2026