Detail of Recall

Type of Product:Device
Medsafe Reference:36649
Brand Name:Hemodynamic Application (MDR)
Model:Philips Interventional Hemodynamic System
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Brand Name:IntelliVue Patient Monitors
Model:Affected systems are Philips Interventional Hemodynamic Systems with Philips Hemodynamic Application software version 1.4.1 or lower
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medizin Systeme Boblingen GmbH
Issue:The Philips Interventional Hemodynamic System automatically deletes the oldest case data when storage is full, but it does not remove export files, snapshots, or generated reports. These accumulate over time and can eventually fill the Hemo Host PC disk. When the disk is full, new cases cannot be opened - the application launches in review mode only. Live signals from the IntelliVue X3 will still display, but users cannot interact with the data. If the disk becomes full during an active case, data will continue to display and functionality remains, but the case will not be saved and cannot be reviewed once closed.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/3/2026