Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 36648 |
| Brand Name: | Syringe Pumps_BD SwissCo |
| Model: | BD Alaris™ neXus and Alaris™ Plus syringe pumps |
| Affected: | Catalog number - 8002TIG03, 8002TIG03-G, 8003TIG03, 8003TIG03-G, 8005TIG03, CCneXus1, CCneXus1-S, PKneXus1 - All serial numbers |
| Software version: | |
| Recalling Organisation: | Becton Dickinson Ltd Unit 4 / 110 Mays Road Onehunga Auckland 1061 |
| Contact Information: | 0800 572 468 |
| Manufacturer: | BD Switzerland Sarl |
| Issue: | The manufacturer has identified through an internal investigation that the BD Alaris™ neXus and Alaris™ Plus syringe pumps may develop a mechanical anomaly on the linear potentiometer if the device is not stored in the correct position. The syringe plunger moves along the linear potentiometer component during infusion. This is an internal component and not readily visible to the Clinical User. If this anomaly occurs, it can lead to incorrect sensor readings, which may trigger error codes PL3, PL8, or PL9. These PL error codes are an intentional design feature that notify the Clinical User that the syringe pump has detected an internal malfunction and can occur for a number of reasons. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/3/2026 |