Detail of Recall

Type of Product:Device
Medsafe Reference:36648
Brand Name:Syringe Pumps_BD SwissCo
Model:BD Alaris™ neXus and Alaris™ Plus syringe pumps
Affected:Catalog number - 8002TIG03, 8002TIG03-G, 8003TIG03, 8003TIG03-G, 8005TIG03, CCneXus1, CCneXus1-S, PKneXus1 - All serial numbers
Software version:
Recalling Organisation:Becton Dickinson Ltd
Unit 4 / 110 Mays Road
Onehunga
Auckland 1061
Contact Information:0800 572 468
Manufacturer:BD Switzerland Sarl
Issue:The manufacturer has identified through an internal investigation that the BD Alaris™ neXus and Alaris™ Plus syringe pumps may develop a mechanical anomaly on the linear potentiometer if the device is not stored in the correct position. The syringe plunger moves along the linear potentiometer component during infusion.
This is an internal component and not readily visible to the Clinical User. If this anomaly occurs, it can lead to incorrect sensor readings, which may trigger error codes PL3, PL8, or PL9.
These PL error codes are an intentional design feature that notify the Clinical User that the syringe pump has detected an internal malfunction and can occur for a number of reasons.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/3/2026