Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:36643
Brand Name:Cook Inc Intraosseous needle
Model:Spectrum 3 Lumen CVC Set PI 7fr 15cm+HC
Affected:Affected lots - 16625683, 16535286, 16584953, 16636220
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Cook Group Inc
Issue:Investigation by the manufacturer of vascular access product lines for Spectrum Vascular, determined that there were errors in the process which sets a product's expiration date based on the earliest expiring component for certain products, rather than a fixed shelf life. These errors resulted in label expiration dates that did not accurately account for all components or materials. As a result, some products were labelled with expiration dates that were longer than the actual shelf life.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/3/2026
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /