Detail of Recall

Type of Product:Device
Medsafe Reference:36641
Brand Name: Efficia DFM100, HeartStart Intrepid
Model:Efficia DFM100 devices with software revision 2.00.33 installed
Affected:Boot-Up Failure with SW 2.00.33
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Goldway (Shenzhen) Industrial Inc
Issue:The manufacturer has identified a potential safety issue with the Efficia DFM100 Defibrillator/Monitor running software version 2.00.33. Affected devices may fail to complete their boot-up sequence. This can occur either: during manual boot-up when the knob is turned to AED, Manual, or Pacer mode, or during a self-test when the device is powered off and connected to any power source. Instead of fully booting, the device may repeatedly display the splash screen and will not become available for clinical use. This condition is indicated to the user by a solid red X shown in the Ready For Use (RFU) indicator.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/3/2026