Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 36611 |
| Brand Name: | Flexicare - Non-Rebreathing Mask |
| Model: | 032-10-075NCNZ Flexicare Adult Non-Rebreathing Green Tinted Nose Clip Mask Head Strap Box/50 (26219001080) |
| Affected: | All Lot numbers |
| Software version: | |
| Recalling Organisation: | Jackson Allison Medical and Surgical Ltd 56 Lunn Ave Mt Wellington Auckland |
| Contact Information: | 09 281 5203 |
| Manufacturer: | Flexicare Medical Ltd |
| Issue: | The affected device has been incorrectly labelled as a non-rebreathing mask. In practice, the mask functions as a rebreathing mask, as it does not include the required visible one-way valves over the side exhalation ports. These valves are essential components that prevent re-inhalation of exhaled gases in a true non-rebreathing system. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 3/3/2026 |