Detail of Recall

Type of Product:Device
Medsafe Reference:36608
Brand Name:VITEK 2 AST-YS08 test kits
Model:VITEK 2 AST-YS08 test kits manufactured on or after 18Jul2025.
Affected:
Software version:
Recalling Organisation:Medi'Ray New Zealand Ltd
53-55 Paul Matthews Road
Albany
Auckland 0632
Contact Information:0800 633 472
Manufacturer:BioMerieux Inc
Issue:Manufacturing Quality Control (QC) failures were observed during release testing for VITEK 2 AST-YS08 test kits manufactured on or after 18Jul2025. Quality Control MIC results for Amphotericin B were out-of-range-high (OORH). To be noted, there are no S/I/R interpretations for Amphotericn B. The antibiotic is reported as MIC only.
If/When using AST-YS08 test kits manufactured on or after 18Jul2025, implement a VITEK 2 Systems BIOART rule to suppress analysis of the Amphotericin B antibiotic, and perform Amphotericin B testing via an alternate method if treatment with Amphotericin B is considered for the patient.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/3/2026