Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 36605 |
| Brand Name: | VITROS Chemistry Products Mg Slides |
| Model: | Catalog number - 825 5093 - 60-Slide Cartridges 192 1204 - 18-Slide Cartridges |
| Affected: | All in-date and future lots |
| Software version: | |
| Recalling Organisation: | Ortho-Clinical Diagnostics c/o Quigg Partners Level 7, 36 Brandon Street Wellington 6011 |
| Contact Information: | 0800 402 899 |
| Manufacturer: | Ortho-Clinical Diagnostics Inc |
| Issue: | The manufacturer became aware of the problem through customer complaints. The VITROS Mg Slides Instructions For Use (IFU) states an On-analyzer stability of = 2 weeks, when the VITROS System is turned on; however, investigation has determined that VITROS Mg Slides results may experience a negative drift outside acceptable limits (greater than 2 within-lab Standard Deviations (WLSD)) sooner than 2 weeks, when cartridges are stored on the VITROS System. This problem affects all current in-date and future lots of VITROS Mg Slides. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/3/2026 |