Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 36604 |
| Brand Name: | Instinct Plus Endoscopic Hemoclip |
| Model: | Instinct® Plus Endoscopic Clipping Device - Product code - G58010 |
| Affected: | All devices manufactured between 2023-02-09 (09 Feb 2023) through to 2025-04-20 (20 Apr 2025) - Multiple lots |
| Software version: | |
| Recalling Organisation: | Obex Medical Ltd Level 3, 109 Carlton Gore Road Newmarket Auckland 1023 |
| Contact Information: | 0800 656 539 |
| Manufacturer: | Wilson-Cook Medical Inc |
| Issue: | The manufacturer is recalling specific lots of Instinct® Plus Endoscopic Clipping Device following reports of malfunction that occurs when the user attempts to open the clip jaws by actuating the handle. The clip housing may detach from the catheter attachment and the clip remaining attached to the drive wire. Please see Figure 1 for a visual representation of the malfunction. The clip will extend from the catheter, but it will remain attached to the internal drive wire instead of opening the clip jaws. The clip cannot be opened if this occurs. The affected devices were manufactured prior to implementation of updated manufacturing processes, which are intended to reduce occurrences of a device malfunction. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/3/2026 |