Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:36590
Brand Name:Soltive System
Model:Olympus SOLTIVE SuperPulsed Laser System (SOLTIVE Laser) products, models Pro (TFL-SLS) and Premium (TFL-PLS)
Affected:Models/SKU - TFL-SLS/ R5000769, TFL-PLS/ R5000768, TFL-CPLU, TFL-CSLU
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Gyrus ACMI Inc
Issue:The manufacturer has received complaints describing an abrupt loss of power and functionality of the SOLTIVE SuperPulsed Laser System, or an inability to power on the unit. The investigation determined that a potential defect in the 24V power supply module in some SOLTIVE Laser units may cause the system to become inoperable. In some cases smoke or a burning smell may also occur. By design, the issue causing any smoke or burning smell that is detected would be contained within the internal laser console enclosure and would be self-extinguishing.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/2/2026
Hide menus