Detail of Recall

Type of Product:Device
Medsafe Reference:36576
Brand Name:Trilogy EVO
Model:Trilogy Evo, Trilogy Evo O2
Affected:Update to the User Manual addendum
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Respironics Inc
Issue:Analysis of the impact of nebulisers on performance of Trilogy Evo Platform ventilators revealed that using non-pneumatic nebulisers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.
Non-pneumatic nebulisers change the airflow characteristics by adding liquid droplets, causing the ventilator's leak estimation calculation to be incorrect. This leads to under-delivery of therapy to the patient even though the graphic user interface does not display the change. This condition may occur with or without aerosol accumulation on the flow sensor. It is also not consistent with the requirements of ISO 80601-2-121.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/2/2026