Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:36575
Brand Name:Trilogy EVO
Model:Trilogy Evo, Trilogy Evo O2
Affected:Update to latest software version 1.06.15.00.
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Respironics Inc
Brand Name:Trilogy EV300
Model:Trilogy EV300
Affected:Update to latest software version 1.06.15.00.
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Respironics Inc
Issue:The manufacturer has become aware of a potential problem when using in-line nebulisers in certain configurations with Trilogy Evo, Trilogy Evo O2, and Trilogy EV300 devices.
The use of in-line nebulisers placed in certain locations can result in aerosol deposits accumulating over time on the device's internal flow sensor.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/2/2026