Detail of Recall

Type of Product:Device
Medsafe Reference:36538
Brand Name:Hugo™ Robotically Assisted Surgery (RAS) System
Model:Bipolar Fenestrated Graspers
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Covidien LLC
Brand Name:Surgical Robotics - Bipolar electrosurgical device
Model:Bipolar Maryland Forceps
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Covidien LLC
Issue:Hugo™ robotic-assisted surgery (RAS) system Bipolar Fenestrated Graspers and Bipolar Maryland Forceps, Customer Facing Numbers MRASI0004 and MRASI0005 respectively, have been identified to have an increased likelihood of component failure within the instrument's drive cable pulley assembly due to manufacturing variations. When this failure occurs, the instrument jaw's actuation and articulation will be incorrect, uncontrolled motion may occur, and a visible small white plastic portion of the instrument's pulley component may become disengaged and loose within the patient's cavity. The Hugo™ RAS System will detect this failure and halt telerobotic control of the affected instrument.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/2/2026