Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:36521
Brand Name:D Dimer
Model:
Affected:
Software version:
Recalling Organisation:Diagnostica Stago Pty Ltd
Level 3
64 Cashel Street
Christchurch
Contact Information:+64 3 377 4115
Manufacturer:Diagnostica Stago SAS
Issue:After receiving customer complaints, the manufacturer conducted thorough investigations and confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range.
This bias affects the specificity values reported in the test's package insert. The recalculated specificity is 63.9% for Pulmonary Embolism (compared to 75.5% stated in the package insert) and 41.2% for Deep Vein Thrombosis (compared to 55.2% stated in the package insert).
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/2/2026