Detail of Recall

Type of Product:Device
Medsafe Reference:36511
Brand Name:STA Liatest Free PS
Model:
Affected:Lot number - 271 971, 272 304, 272 529, 272 859, 273 018, 273 426, 273 635
Software version:
Recalling Organisation:Diagnostica Stago Pty Ltd
Level 3
64 Cashel Street
Christchurch
Contact Information:+64 3 377 4115
Manufacturer:Diagnostica Stago SAS
Issue:The manufacturer identified out-of-range control results and an underestimation of free protein S levels in normal patient samples associated with STA Liatest Free Protein S 6 (Lots 271971, 272304, 272529, 272859, 273018, 273426, 273635). The issue affects only results within the normal-range and was traced to a specific lot of a biological raw material component, the impact of which became progressively apparent over time.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/1/2026