Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 36511 |
| Brand Name: | STA Liatest Free PS |
| Model: | |
| Affected: | Lot number - 271 971, 272 304, 272 529, 272 859, 273 018, 273 426, 273 635 |
| Software version: | |
| Recalling Organisation: | Diagnostica Stago Pty Ltd Level 3 64 Cashel Street Christchurch |
| Contact Information: | +64 3 377 4115 |
| Manufacturer: | Diagnostica Stago SAS |
| Issue: | The manufacturer identified out-of-range control results and an underestimation of free protein S levels in normal patient samples associated with STA Liatest Free Protein S 6 (Lots 271971, 272304, 272529, 272859, 273018, 273426, 273635). The issue affects only results within the normal-range and was traced to a specific lot of a biological raw material component, the impact of which became progressively apparent over time. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/1/2026 |