Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 36504 |
| Brand Name: | Infusion pump administration set |
| Model: | Alaris™ Products. Alaris™ VP Volumetric Pump. Alaris™ Safety Clamp. INFUSION SET. 15 µm FILTER |
| Affected: | Product code - 70593 lot number 1032633 |
| Software version: | |
| Recalling Organisation: | Becton Dickinson Ltd Unit 4 / 110 Mays Road Onehunga Auckland 1061 |
| Contact Information: | 0800 572 468 |
| Manufacturer: | BD Switzerland Sarl |
| Issue: | The manufacturer has identified, as a result of customer complaints received, that the specific product code and lot number of Alaris™ VP Volumetric Pump Infusion Sets have the potential for fluid leakage from the silicone tube at the end of the pumping segment. Leakage at the silicone pumping segment during priming or infusion may result in under-infusion of prescribed therapy, unintended exposure to hazardous drugs, and increased risk of air embolism or bacterial contamination if urgent set replacement compromises sterility. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 27/1/2026 |