Detail of Recall

Type of Product:Device
Medsafe Reference:36497
Brand Name:Olympus - Insufflator, endoscopic (Japan)
Model:Unit models UHI, UHI-2, and UHI-3
Affected:All affected numbers -Olympus High Flow Insufflation Unit models - UHI, UHI-2, UHI-3 and N1000161
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Brand Name:UHI-4_High Flow Insufflator Unit
Model:Use available alternatives including an existing UHI-4 if available
Affected:
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Issue:The manufacturer has determined that the software algorithm on High Flow Insufflation Unit, models UHI, UHI-2, and UHI-3, requires correction to address a potential issue which could lead to overpressure events. These events may occur due to an over insufflation of the abdominal cavity resulting from use of the UHI, UHI-2, or UHI-3 during patient procedures. This includes events where the device may not alarm or otherwise notify the user and/or may not relieve the over insufflation to the set pressure. Because no corrective solution is available, the manufacturer has decided to remove these devices from the market.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/1/2026