Detail of Recall

Type of Product:Device
Medsafe Reference:36485
Brand Name:The Venous Bubble Sensor
Model:BS 3/8x3/32 L1.7 Mat. 701055720 (Bubble Sensor for 3/8 x 3/32 tubing, length 1,7 m)
Affected:
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Maquet Cardio Pulmonary GmbH
Issue:The manufacturer has identified a problem with the durability of the connecting cable near the connection to the Bubble Sensor. Excessive bending of the cable can lead to poor contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device (ROTAFLOW II or CARDIOHELP-i). These errors can occur temporarily when the cable is moved or permanently if the connection is fully compromised.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/1/2026