Detail of Recall

Type of Product:Device
Medsafe Reference:36483
Brand Name:Ultrasound Systems
Model:Lumify Systems incompatible with iOS 18 devices
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Ultrasound LLC
Issue:The manufacturer has identified that the Lumify software application has an intentional design component that restricts the ability to scan if the host device is charging. The Lumify Power Module (LPM) provides battery power to support the use of the transducer. It is not intended to provide charging capability to iOS devices. With version 18.X of iOS, when the device's charge level drops below 99%, it will attempt to charge from the LPM through the USB connection. This does not happen for Lightning port connected devices. If the battery threshold is crossed while live scanning, the application will interrupt the live scan and display a dialog message requiring the user to charge the tablet. If the battery level is below the threshold prior to starting live scanning the application will prevent live scanning from starting and display a dialog message requiring the user to charge the tablet.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/1/2026