Detail of Recall

Type of Product:Device
Medsafe Reference:36480
Brand Name: Philips MR systems Ingenia, Achieva Series
Model:Philips MR systems with SW version R11.1 and R12.1
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has identified potential stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture
Archiving and Communication System (PACS).
The MRE viewer PACS indicates around 2 kPa (incorrect stiffness values), which can be associated with healthy tissue, while the general viewing environment in the MR Console viewer indicates almost 8 kPa (correct stiffness values), which can be an indicator of cirrhosis.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/1/2026