Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:36470
Brand Name:Olympus - Sheaths Laser Sidefire
Model:SHEATH LASER SIDEFIRE OUTER 22.5Fr
Affected:All serial numbers
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Winter & Ibe GmbH
Issue:In October 2024, Olympus issued a Product Correction in relation to the Cystoscope Outer sheath, 22.5 Fr (Medsafe Ref #33877). The Product Correction Notice instructed users that the Cystoscope Outer sheath should not be used with a GreenLight Laser for BPH therapy until further notice. The manufacturer has received reports about a Cystoscope Outer sheath damaged tip during use of a laser probe. The manufacturer conducted additional testing, which indicated further regulatory actions would be required in order to state compatibility of the cystoscopy sheath with GreenLight laser fibres. Accordingly, and have made the decision to discontinue the product and remove the device from the market.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/1/2026