Detail of Recall

Type of Product:Device
Medsafe Reference:36468
Brand Name:Omni Legend System Images
Model:Omni Legend PET/CT, all configurations using Software version SP5,SP6,SP7,SP8
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems Functional Imaging Haifa
Issue:The manufacturer has become aware of a potential intermittent issue on certain Omni Legend systems that can result in a streaking artifact in the PET clinical scan images. In some cases, the artifact can mimic pathology and has the potential to lead to misdiagnosis if undetected. (Note: Standalone CT image acquisition is not impacted by this issue.)
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/1/2026