Detail of Recall

Type of Product:Device
Medsafe Reference:36467
Brand Name:Endoscopic electrosurgical handpiece
Model:Model: KD-V411M & KD-V431M - CleverCut 3V Single Use 3-Lumen Sphincterotome V
Affected:Multiple lots
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Brand Name:Andotherapy device, active, single-use
Model:Model: KD-VC411Q & KD-VC431Q - CleverCut 3V Single Use Sphincterotome V (Distal Wireguided) with C-Channel
Affected:Multiple lots
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Medical Systems Corporation
Issue:The manufacturer has observed an increase in complaints regarding wire deformation in the CleverCut 3V Sphincterotome devices. Preliminary investigations showed that units manufactured with a thermoforming process, which involves heating the assembly, so the device tip keeps the curved stylet's shape, are less likely to have an incorrect cutting wire orientation. Devices which did not undergo thermoforming could deform and lose performance. As a result, the manufacturer is removing lots where the devices may not have been manufactured with the thermoforming process.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/1/2026