Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:36459
Brand Name:Philips HEARTSTREAM Efficia DFM100 Defibrillator/Monitor
Model:HEARTSTREAM Efficia DFM100 Defibrillator/Monitor
Affected:Efficia DFM100 devices with software revision 2.00.33 installed
Software version:
Recalling Organisation:Clinical & Regulatory Services Limited
30 Fulford Road
Havelock North
Hastings 4130
Contact Information:(06) 844 4490
Manufacturer:Philips Goldway (Shenzhen) Industrial Inc
Issue:The Efficia DFM100 Defibrillator/Monitor, with software revision 2.00.33 installed, may fail to complete its boot-up sequence. This could occur either 1) during manual boot-up if the knob is turned to AED/Manual/Pacer or 2) as part of a self-test if the device is turned off and connected to any power source. Instead of completing the boot-up sequence, the device may repeatedly display the splash screen and will not be available for clinical use. This will be indicated to the user by displaying a solid red X in the Ready For Use (RFU) indicator.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/1/2026