Detail of Recall

Type of Product:Device
Medsafe Reference:36440
Brand Name:ShockPulse Lithotripsy System
Model:ShockPulse-SE Lithotripsy System – Reusable Probes and ShockPulse Lithotripsy Generator
Affected:Serial number - CG6069
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Gyrus ACMI Inc
Issue:The manufacturer has identified through an internal investigation that specific serial numbers of ShockPulse generators may have a mis-wired component. This condition does not impact the intended functionality of the device; however, the miswiring may introduce additional electrical noise on the power supply output. The presence of noise on the ultrasonic input circuit is not compliant with applicable electromagnetic compatibility (EMC) standards and may adversely affect the overall reliability of the system.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/1/2026