Detail of Recall

Type of Product:Device
Medsafe Reference:36438
Brand Name:Resectoscope Resection Sheaths
Model:RESECTION SHEATH and INNER SHEATH
Affected:Multiple lots
Software version:
Recalling Organisation:Olympus (NZ) Ltd
15D Paul Matthews Road
Albany
Auckland
Contact Information:0508 659 6787
Manufacturer:Olympus Winter & Ibe GmbH
Issue:The manufacturer has identified complaints related to the ceramic tip of the resection/inner sheath breaking. There were reports, where the tip fell off during the procedure. In most events, the clinician identified and retrieved the fragment during the procedure. By nature, ceramics are not pliable and may fracture unexpectedly under significant mechanical impact. The manufacturer' investigation revealed that some devices remain in use for prolonged periods and may be subjected to stress over their operational lifetime.
The manufacturer is highlighting the importance of strictly adhering to the 'Inspection and Testing' and 'Caution' sections in the IFU for the resection/inner sheath, as well as the 'Inspection Before Each Use' section in the Resectoscope System Manual.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/1/2026