Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 36429 |
| Brand Name: | Ventilator circuit for non-invasive ventilation |
| Model: | Product Code - AMVC1792-156 |
| Affected: | Lot numbers 021025, 181024 |
| Software version: | |
| Recalling Organisation: | Active Rehab Equipment 25 Rennie Drive Airport Oaks Auckland |
| Contact Information: | 0800 336 339 |
| Manufacturer: | Armstrong Medical Ltd |
| Issue: | The manufacturer investigation of the customer reports did not identify any anomaly or non-conformance with the product as all devices were manufactured to the specifications set. The existing IFU requires an update to prevent reasonable forseeable misuse when handling the device. Pulling on the tubing may lead to inadvertent tubing separation from the purple cuff connector. Replacement of the circuit may result in delays in treatment and will require intervention by Healthcare Professional. These situations present a potential risk of respiratory distress, hypoxia, and/or low oxygen saturation as the Healthcare Provider swaps out the breathing circuit. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 16/2/2026 |