Detail of Recall

Type of Product:Device
Medsafe Reference:36429
Brand Name:Ventilator circuit for non-invasive ventilation
Model:Product Code - AMVC1792-156
Affected:Lot numbers 021025, 181024
Software version:
Recalling Organisation:Active Rehab Equipment
25 Rennie Drive
Airport Oaks
Auckland
Contact Information:0800 336 339
Manufacturer:Armstrong Medical Ltd
Issue:The manufacturer investigation of the customer reports did not identify any anomaly or non-conformance with the product as all devices were manufactured to the specifications set. The existing IFU requires an update to prevent reasonable forseeable misuse when handling the device.
Pulling on the tubing may lead to inadvertent tubing separation from the purple cuff connector. Replacement of the circuit may result in delays in treatment and will require intervention by Healthcare Professional. These situations present a potential risk of respiratory distress, hypoxia, and/or low oxygen saturation as the Healthcare Provider swaps out the breathing circuit.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/2/2026