Detail of Recall

Type of Product:Device
Medsafe Reference:38060
Brand Name:Hamilton/47244/Ventilators
Model:HAMILTON-C6 Part number: 160021
Affected:Serial numbers: 17196, 17357, 17358, 17361, 17362, 17363
Software version:
Recalling Organisation:InterMed Medical Ltd
71 Apollo Drive
Albany
AUCKLAND 1311
Contact Information:0800 333 444
Manufacturer:Hamilton Medical AG
Issue:Certain ventilators with serial numbers are equipped with a blower module that is prone to reach the end of its service life earlier than expected. In some cases, the ventilator was alarming with technical events or technical faults (e.g. TE 231009 blower speed low or TF 431001 blower fault), indicating a technical fault of the blower. The issue might occur even if the pre-operative test of the ventilator passed.
If the ventilator alarms with a technical fault during active ventilation with a patient, the device might transition to ambient state, resulting in loss of mechanical ventilation and requiring immediate provision of alternative ventilation.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/11/2026