Database updated: 27 November 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:38058
Brand Name:Corpuls 3 Defib including Accessories and Upgrades
Model:Corpuls 3 and Corpuls 3T
Affected:Product code: CP-04100, CP-04101 Software versions: 4.3.X and 4.4.X
Software version:
Recalling Organisation:Universal Specialities Ltd trading as USL Medical
494 Rosebank Road
Avondale
AUCKLAND 1026
Contact Information:09 829 0595
Manufacturer:GS Elektromedizinische GmbH
Issue:Two safety concerns have been identified through post market monitoring and surveillance: ongoing concerns related to shock performance and reboot behaviour. Reports of no shock delivered (inclusive of events of failure to shock, shock aborted, defibrillation failure and defibrillation alarms) continued to be received.
The manufacturer has been made aware of reboots in devices with software version 4.4.4/4.4.5, which was implemented to correct reboots caused by certain software bugs. Under certain conditions, some functions are temporarily paused in affected devices while the device spontaneously reboots.
Users are also being notified of four additional problems that are under active investigation: end-tidal CO2 measurement inaccuracies, ECG transmission delays, failure to recognise STEMI and heart rate discrepancies.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/11/2026
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /