Detail of Recall

Type of Product:Device
Medsafe Reference:38039
Brand Name:Corpuls 3 Defib including Accessories and Upgrades
Model:Corpuls 3 and Corpuls 3T
Affected:Product Code: CP-04100, CP-04101 - Multiple serial numbers
Software version:
Recalling Organisation:Universal Specialities Ltd trading as USL Medical
494 Rosebank Road
Avondale
AUCKLAND 1026
Contact Information:09 829 0595
Manufacturer:GS Elektromedizinische GmbH
Issue:The manufactuere has identified a potential correlation between printing a resting ECG and a temporary interruption of the monitor's Bluetooth connection, which may result in an unexpected, brief interruption of the display.
A temporary interruption of the display may result in users temporarily losing access to displayed patient information and being unable to operate the system for 5–25 seconds. Particularly in emergency situations, this may lead to an interruption or delay in life-saving treatment and, in the worst case, may result in serious adverse health consequences for patients.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/11/2026