Database updated: 27 November 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 38022 |
| Brand Name: | G3R vital signs monitor with mobile stand |
| Model: | Model: G3R. G3R Vital Signs Monitor – Non-compliant Power Cord |
| Affected: | Product code: 4604 |
| Software version: | |
| Recalling Organisation: | Cubro Ltd 3 Hynds Rd Tauranga |
| Contact Information: | 0800 656 527 |
| Manufacturer: | Guangdong Transtek Medical Electronics Co Ltd |
| Issue: | The manufacturer has identified that certain New Zealand mains power cords supplied with the G3R Vital Signs Monitor do not comply with AS/NZS 3112 because the plugs do not have the required insulation on the active and neutral pins. This presents a potential electric shock hazard if a person comes into contact with a live pin while inserting or removing the plug from a power outlet. Electric shock may result in injury. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 19/11/2026 |