Detail of Recall

Type of Product:Device
Medsafe Reference:38022
Brand Name:G3R vital signs monitor with mobile stand
Model:Model: G3R. G3R Vital Signs Monitor – Non-compliant Power Cord
Affected:Product code: 4604
Software version:
Recalling Organisation:Cubro Ltd
3 Hynds Rd
Tauranga
Contact Information:0800 656 527
Manufacturer:Guangdong Transtek Medical Electronics Co Ltd
Issue:The manufacturer has identified that certain New Zealand mains power cords supplied with the G3R Vital Signs Monitor do not comply with AS/NZS 3112 because the plugs do not have the required insulation on the active and neutral pins.
This presents a potential electric shock hazard if a person comes into contact with a live pin while inserting or removing the plug from a power outlet. Electric shock may result in injury.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/11/2026