Detail of Recall

Type of Product:Device
Medsafe Reference:37838
Brand Name:Liquichek Unassayed Chemistry Control
Model:
Affected:Lot Number: 93031, 93041, 93032, 93042, 93030, 93040
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Brand Name:Liquid Unassayed Multiqual
Model:
Affected:Lot Numbers: 56761, 56762, 56763, 56760
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Brand Name:Liquid Assayed Multiqual
Model:
Affected:Lot Numbers: 46021, 46041, 46022, 46042, 46023, 46043, 46020, 46040
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Brand Name:Liquid Assayed Multiqual Premium
Model:
Affected:Lot Numbers: 86991, 86992, 86993, 86990
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Issue:The manufacturer has identified a stability issue affecting the quality control material lots listed above. Investigation found that Triglycerides, Phosphorus, and Magnesium values increase over time after the vial is thawed and opened (open-vial stability). Magnesium values also decrease over the product's shelf life.
Use of affected quality control lots may result in QC shifts for the impacted analytes, which could require laboratory investigation, repeat testing, or use of alternate control material with subsequent temporary operational delays related to QC investigation.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/11/2026