Database updated: 20 November 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:37768
Brand Name:Access HIV Ag/Ab Combo Calibrator
Model:Ref #C39451
Affected:Multiple lots
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The manufacturer has determined through internal investigation that the calibrator card 2D barcode for the Access HIV Ag/Ab combo calibrator contain reversed assay calibrator value assignments. As a result, the DxI 9000 Access Immunoassay Analyzer attributes the HIV-Ag calibrator assigned values to the HIV-Ab assay, and the HIV-Ab calibrator assigned values to the HIV-Ag assay.
Access HIV Ag/Ab combo Calibrator may generate false Nonreactive or Reactive Access HIV Ag/Ab combo result interpretations.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/10/2026
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /