Detail of Recall

Type of Product:Device
Medsafe Reference:37756
Brand Name:MRI system, full-body, superconducting magnet
Model:SIGNA Artist Evo and SIGNA Artist electronics cabinet
Affected:GTIN Number - 00195278117021, 00195278481382
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Healthcare Tianjin Company Ltd
Issue:The manufacturer has become aware of an exposed sharp edge inside the SIGNA Artist Evo and SIGNA Artist electronics cabinet that could damage power cables, which may potentially cause an electrical fault. If the electrical fault is present and the cabinet door is opened while the system is powered, this could potentially result in an arc flash that could lead to serious injury to service personnel in immediate proximity.
The issue would only present itself during servicing activities involving the electronics cabinet. The electronics cabinet is located within the equipment room and should only be accessed by qualified service personnel.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/10/2026