Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37756 |
| Brand Name: | MRI system, full-body, superconducting magnet |
| Model: | SIGNA Artist Evo and SIGNA Artist electronics cabinet |
| Affected: | GTIN Number - 00195278117021, 00195278481382 |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Healthcare Tianjin Company Ltd |
| Issue: | The manufacturer has become aware of an exposed sharp edge inside the SIGNA Artist Evo and SIGNA Artist electronics cabinet that could damage power cables, which may potentially cause an electrical fault. If the electrical fault is present and the cabinet door is opened while the system is powered, this could potentially result in an arc flash that could lead to serious injury to service personnel in immediate proximity. The issue would only present itself during servicing activities involving the electronics cabinet. The electronics cabinet is located within the equipment room and should only be accessed by qualified service personnel. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 21/10/2026 |