Detail of Recall

Type of Product:Device
Medsafe Reference:37737
Brand Name:[MDR] Philips Lumify Topical Transducer
Model:Lumify Diagnostic Ultrasound System
Affected:All Software is impacted: Android Software versions: 1.X, 3.X, 4.X, 5.X iOS Software Versions: 2.X and 5.X
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Ultrasound LLC
Brand Name:[MDR]Philips Lumify Ultrasound Systems
Model:Lumify Diagnostic Ultrasound System
Affected:All Software is impacted: Android Software versions: 1.X, 3.X, 4.X, 5.X iOS Software Versions: 2.X and 5.X
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Ultrasound LLC
Issue:The manufacturer has identified the following potential problems with the Lumify Diagnostic Ultrasound System which may prevent or delay scanning functionality:
1. Due to a software problem, the Lumify application may incorrectly require a previously registered transducer to be re-registered before use. Scanning cannot proceed until registration is completed, which requires Wi-Fi or cellular connectivity. In environments without network access, such as ambulances, aircraft, or remote settings, this may delay patient care.
2. The connection between the Lumify transducer USB cable and the customer's tablet or phone may become unstable or disconnect during use. This can interrupt scanning and device functionality, potentially resulting in a delay in patient care.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/10/2026