Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37702 |
| Brand Name: | Cooper - Apple Hunts |
| Model: | 5MM APPLE HUNT SECONDARY CANNULA PYRAMIDAL TROCAR |
| Affected: | Product code: CO-900-800 - Affected lot number: 550034148 |
| Software version: | |
| Recalling Organisation: | Obex Medical Ltd Level 3, 109 Carlton Gore Road Newmarket Auckland 1023 |
| Contact Information: | 0800 656 539 |
| Manufacturer: | CooperSurgical Inc |
| Issue: | The manufacturer is recalling a specific lot of the Apple-Hunt Secondary Cannula™ Pyramidal Tip (5mm) following reports of devices that may contain a cannula component that is longer than the intended specification. As a result, when the device is assembled in accordance with the Instructions for Use, the pyramidal trocar tip does not extend beyond the distal end of the cannula as intended. The condition is readily detectable during the required assembly and pre-use inspection process, as the trocar tip remains enclosed within the cannula following assembly. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/10/2026 |