Detail of Recall

Type of Product:Device
Medsafe Reference:37695
Brand Name:[MDR] Radiography 7000 M
Model:Radiography 7000M with Software release R1.0.3
Affected:Product number: 712010
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:The manufacturer has identified that when the operator is switching between examinations, an intermittent software failure may cause the console to display the message "Call service! Filter selection failure. This display message may appear even if the motorized filter has been selected correctly. If this happens, the system enters a Not Ready state and does not permit an X-ray exposure.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/10/2026