Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 37648 |
| Brand Name: | Intelerad - InteleradPACS |
| Model: | Multiple versions |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Clinical & Regulatory Services Limited 30 Fulford Road Havelock North Hastings 4130 |
| Contact Information: | (06) 844 4490 |
| Manufacturer: | Intelerad Medical Systems Incorporated |
| Issue: | The manufacturer has become aware of two problems affecting certain versions of IntelePACS after identifying two separate issues. Problem 1: A security vulnerability in the InteleBrowser administration portal could allow an authenticated user to gain unauthorised administrator access, potentially resulting in data modification, deletion, or restricted system access. Problem 2: A software issue could result in unintended image data deletion during IOCM processing in systems with a specific non-default cloud archive configuration. Intelerad has found no evidence of this configuration being used in any production environment. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/9/2026 |