Detail of Recall

Type of Product:Device
Medsafe Reference:37646
Brand Name:ARTIS pheno, ARTIS icono floor, ARTIS icono biplane, ARTIS icono ceiling with software version VE50B
Model:Catalogue/Lot number – ARTIS icono biplane - 11327600 with SW VE50B; ARTIS icono ceiling - 11328100 with SW VE50B; ARTIS icono floor - 11327700 with SW VE50B; ARTIS pheno - 10849000 with SW VE50B
Affected:With software VE50B
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthineers AG
Issue:The manufacturer has identified that in certain operating situations, the system may fail to execute movement control commands as intended. If this occurs, motorised stand and/or table movement may no longer be available until the system has been completely restarted. The impact on system operation is that movement of the affected system axes may no longer be possible until the system is restarted.
This may limit or delay the ability to change projection planes or to position the system as required for the intended procedure. In situations where immediate access to the patient or repositioning of the system is required, including during cardiopulmonary resuscitation (CPR) activities, the described behaviour may interfere with the intended workflow.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/9/2026