Type of Product:Device
Medsafe Reference:37637
Brand Name:VITEK® 2 AST-XN24
Model:Reference number - 424351
Affected:Drug impacted: cza02n/ipr01n/mev01n, cza02n/ipr01n/mev01n
Software version:
Recalling Organisation:Medi'Ray New Zealand Ltd
53-55 Paul Matthews Road
Albany
Auckland 0632
Contact Information:0800 633 472
Manufacturer:bioMerieux Italia S.p.A.
Issue:The manufacturer has identified an increase in complaints for out-of-range high (OORH) quality control (QC) results for the following antibiotic formulations: Ceftazidime/Avibactam (cza02n), Imipenem/Relebactam (ipr01n), and Meropenem/Vaborbactam (mev01n). During the investigation
we identified that a small number of VITEK® 2 GN-AST cards users, have also reported false resistant antibiotic results with the same antibiotics when testing strains of Enterobacterales/Enterobacteriaceae species and/or Pseudomonas aeruginosa.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory